Trouble-shooting
PolyLink kit
Trouble-shooting
Problem
Possible cause
Suggestion
No acquisition of the
signals of effort /
position.
The PolyLink module is
not present or not
correctly inserted in the
S.Box device.
Check that the PolyLink module is properly
inserted in the accessory location of the
device. Otherwise, reinstall it (see "Insertion of
the PolyLink module in the S.Box device" page
8).
The PolyLink
effort/position kit does
not work.
With the tip of a pen, check that the switch of
the effort/position kit is set to ON.
Turn off, then turn on the effort/position kit
and check that the LED lights in green.
Check the charge of the internal battery. If the
LED is lit in orange or in red, recharge the
internal battery by using the USB charger (see
"Implementation of the effort/position kit and "
page 8).
The thorax and
abdomen effort sensors
are not properly
connected.
Check
the
connections
between
the
effort/position kit, the thorax and abdomen
effort sensors and the belts. Readjust them if
necessary.
If the problem persists, contact your Health
Professional.
The LED of the
effort/position kit is off.
The extinction of the LED is automatic when
the effort/position kit has been working for
1 minute.
If the LED never lights up, contact your Health
Professional.
The symbol
is not
present in the center of
the display when the
S.Box device is working.
The BLE (Bluetooth Low
Energy) communication
is not enabled.
Check that the BLE communication is enabled
in the patient settings menu of the S.Box
device. Otherwise, activate it (BLE:ON). See
"Communication with the PolyLink kit" page 9.
If the problem persists, contact your Health
Professional.
Communication
impossible with the
PolyLink kit.
The PolyLink module is
not paired with the
PolyLink effort/position
kit.
Check that the PolyLink module is paired in the
settings menu of the S.Box device. See
"Communication with the PolyLink kit" page 9.
If the problem persists, contact your Health
Professional.
The PolyLink kit seems
to experience
interference and does
not work properly.
Too high
electromagnetic
disturbances.
Move the S.Box device and the PolyLink kit
away from sources of interference, such as
halogen lamps, cordless phones, etc.