background image

Consultation with SebaKMT 

 

 

User Manual 

Data Logger 

Sebalog D-3 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Issue: 

06 (02/2019) - EN 

Article number: 

83914 

Mess- und Ortungstechnik 
Measuring and Locating Technologies

 

Elektrizitätsnetze 
Power Networks 

 

Kommunikationsnetze 
Communication Networks 

 

Rohrleitungsnetze 
Water Networks

 

 

Abwassernetze 
Sewer Systems 

 

Leitungsortung 
Line Locating 

 

Содержание Sebalog D-3

Страница 1: ...sue 06 02 2019 EN Article number 83914 Mess und Ortungstechnik Measuring and Locating Technologies Elektrizit tsnetze Power Networks Kommunikationsnetze Communication Networks Rohrleitungsnetze Water...

Страница 2: ...Consultation with SebaKMT 2...

Страница 3: ...ervice station please contact Seba Dynatronic Mess und Ortungstechnik GmbH Hagenuk KMT Kabelmesstechnik GmbH Dr Herbert Iann Str 6 D 96148 Baunach Phone 49 9544 68 0 Fax 49 9544 22 73 R deraue 41 D 01...

Страница 4: ...MT are hereby limited to a period of 12 months from the date of delivery Each component supplied by SebaKMT within the context of warranty will also be covered by this warranty for the remaining perio...

Страница 5: ...18 3 4 2 External power supply 18 3 4 3 Checking the battery status 19 3 5 Memory 20 3 6 Available versions 20 4 Data encryption and password 21 4 1 Introduction 21 4 2 Encryption of the radio commun...

Страница 6: ...vices 48 7 7 4 Change password 49 7 7 5 Disable password protection 50 8 Programming the logger using the SebaDataView 3 software 51 8 1 Introduction 51 8 2 Status of the device 52 8 3 Selecting the l...

Страница 7: ...urement data display 74 9 4 3 Displaying pressure surges 76 9 5 Exporting data in CSV format 77 10 Miscellaneous functions of the SebaDataView 3 software 79 10 1 Find out status information of a devic...

Страница 8: ...Terms of Warranty 8...

Страница 9: ...the authorised and trained operator He needs to read the manual thoroughly The manufacturer is not liable for damage to material or humans due to non observance of the instructions and safety advices...

Страница 10: ...This device is designed for industrial use When used at home it could cause interference to other equipment such as the radio or television The interference level from the line complies with the limi...

Страница 11: ...sensor Internal pressure sensor optional 16 bar 25 bar 35 bar accuracy 1 resolution 0 006 bar 16 bar sensor Log interval standard measurement pressure surge meas Standard measurement 1 sec up to 24 h...

Страница 12: ...uch Pressure hose 2 m 35 bar 118304220 GSM Antenne GSM antenna with connection cable 2 m 820020888 The following optional accessory parts can be ordered from SebaKMT Accessory part Description Item No...

Страница 13: ...pressure curve with a narrow measuring interval Using the real time measurement function or the Step Test function a measurement can be carried out and observed live on a computer The logger can be us...

Страница 14: ...rror during ftp data transfer LED turns off when modem was successfully turned down wrong PIN code flashing red 3x before turning on constantly SIM card blocked PUK needed flashing white 3x before tur...

Страница 15: ...socket for connecting an external power supply or alarm triggering devices Connector P1 pressure sensor internal measuring channel Channel 1 GSM socket for connecting the external GSM antenna IN socke...

Страница 16: ...ust have a radio interface For that purpose the Log RI or Log RI can be connected Furthermore if your logger is equipped with an internal GSM module measurement data uploads to an FTP server or config...

Страница 17: ...1x green M switching on blue light M data transfer in progress red light M malfunction We recommend that the device is always operated with the latest firmware More exact information on how to update...

Страница 18: ...r switches itself off Measurement data logged up to that point remain safely in the loggers memory until the next programming Flat batteries can be replaced see page 85 A set of matching spare batteri...

Страница 19: ...nt battery status of the logger is also sent to the receiver In the SebaDataView 3 Software you can find the latest battery status in the device s status information see page 79 To check the current b...

Страница 20: ...code on the logger s label showshow the device is equipped The type code consists of the following segments Due to the large amount of possible configurations this manual cannot state explicitly wheth...

Страница 21: ...above see table in the previous section work with an encryption protocol for wireless communication This means that certain user sensitive data will only be transmitted encrypted These data are subjec...

Страница 22: ...ons who have the password for this user level are considered main users They have the full functionality of the software at their disposal Main users are allowed to manage passwords and extend passwor...

Страница 23: ...m messages or status messages via e mail an e mail account is needed from which the messages are to be sent If the mail server of your company is not capable for this task e g due to a firewall a webm...

Страница 24: ...er on the housing but to take both parts upside down and then put the housing on the cover Note the following points when assembling The rubber seals and the housing and cover surfaces which touch eac...

Страница 25: ...to be taken away from the contact area Afterwards the logger switches off and the lamp goes out Make sure to take the magnet away immediately after the lamp switches to red Otherwise the logger reboo...

Страница 26: ...Commissioning the logger 26...

Страница 27: ...software Please refer to the according chapter in order to get a detailed description of the necessary configuration steps see page 51 External sensors connected cannot be supplied by the data logger...

Страница 28: ...sensitive to overpressure Do not exceed the measuring range in order to avoid irreparable damage to the sensor Do not undercut the specified operating temperature of the pressure sensors in order to...

Страница 29: ...white Channel 1 Channels up to 4 sensors can be connected to brown Channel 1 green Channel 2 yellow Channel 2 grey Channel 3 pink Channel 3 blue Channel 4 red Channel 4 black not used purple not used...

Страница 30: ...he purple wire GND green Switching Input 2 brown Relay 1 OUT An electric circuit with an external DC power supply up to 12 V 1 A can be connected to each switching output If an internal relay is trigg...

Страница 31: ...ensor manual for detailed information The following example describes how to connect a sensor with a 0 5 V voltage output or 0 Hz 6000 Hz frequency output or pulse output The following example describ...

Страница 32: ...of the internal relays are switched in order to activate trigger the load connected to the respective switching output in this case a signal lamp and a SMS and or e mail can be sent to predefined rec...

Страница 33: ...e connection cables Care shall be taken to ensure that cables and their connections are not subjected to tensile load Do not hang the logger on its connection cables Make sure the logger is switched o...

Страница 34: ...Connecting the logger and installing it on site 34...

Страница 35: ...ecute the installation file and follow the instructions on the screen The application is installed to the following folder C Program Files SebaKMT SebaDataView Furthermore a database is created in the...

Страница 36: ...7 and later In all menu levels the display shows the following structure All function and command buttons are arranged in a panel called ribbon as it is known from Microsoft Office applications Every...

Страница 37: ...abase is displayed All the devices added to the database by means of their identification number can be found in this so called directory tree The directory tree shows the following structure of folde...

Страница 38: ...tion 1 Mark the folder to be deleted 2 Open the tab Directory in the multifunction bar 3 In the segment Folder click on Delete 4 Answer the security query with Yes Result The corresponding folder is r...

Страница 39: ...ich the new group should be created 2 Open the tab Directory in the multifunction bar 3 In the segment Group click on New 4 In the window which opens enter the Name and a Comment for the new group and...

Страница 40: ...ftware to which the various devices are added according to their type Example Group A has Log N3 LogDX Log D3 and SebaFlow devices A radio interface e g Log RI needs to be connected to the computer in...

Страница 41: ...s where the newly created map is shown 5 To mark the place of installation of a device in the directory tree click on the device concerned keep the left mouse button pushed and drag the device to the...

Страница 42: ...mage file and now appears in the directory tree in the form of a sub folder called Map A new window opens where the newly created map is shown 7 To mark the place of installation of a device in the di...

Страница 43: ...egment of the menu bar see page 43 To mark the place of installation of a device in the directory tree click on the device concerned keep the left mouse button pushed and drag the device to the point...

Страница 44: ...ave the chance to costumize the map view For this purpose open the Directory tab of the multifunction bar There in the Pipe network segment the following checkboxes are available Option Description 10...

Страница 45: ...ocation segment The current target directory opens in an Explorer window With the default setting it is the data directory mentioned above The precise target path is displayed in the address line You...

Страница 46: ...system administrator or the server operator for this data or it is available in your FTP usage agreement The Email Settings segment contains boxes for entering the access data for the sender email acc...

Страница 47: ...started a login area always appears which requires the user to select the user level and enter the corresponding password Depending on the user level the software then opens with limited or full funct...

Страница 48: ...t From now on a login area appears in the start screen above the language selection To log in proceed as follows Step Description 1 Click on the drop down menu and select a user level 2 Enter the corr...

Страница 49: ...assword protection again to the desired devices Set password User level 1 login for the SebaDataView 3 software is required User levels 2 and 3 are not authorized to manage the passwords To change the...

Страница 50: ...as follows Step Description 1 Start the software and log in to User level 1 2 Click on the water drop symbol in the top left 3 Click the Settings button in the appearing context menu 4 Click the Chan...

Страница 51: ...ial programming also be remotely programmed from the office via an FTP server For that the following requirements must be met computer with SebaDataView 3 software the access data for the FTP server m...

Страница 52: ...on the device You find detailed information about the necessary programming steps in the following sections 8 2 Status of the device In the State segment current status information about the device ca...

Страница 53: ...vailable for selection After a channel has been activated it has to be specified which type of sensor is connected to the channel Click the respective Configure button A new window appears 8 4 1 Selec...

Страница 54: ...specify how much a measuring value has to differ from the previous value in order to classify a pressure fluctuation as a pressure surge Example If 0 5 is set pressure surge recordings start as soon a...

Страница 55: ...ed select the option User defined A new input field opens Enter the desired measuring unit there 3 Depending on which input type or which unit was selected additional input fields are displayed Assign...

Страница 56: ...ar correlates with 5 V As a result the fields must be populated as follows NOTE The maximum permissable input voltage is 5 V A level meter with frequency output is installed in a 430 l tank Input type...

Страница 57: ...the chance to enter the water meters counter reading into the respective field prior to the measurement A flowmeter is connected to channel 3 In addition to the flow rate itself the flow direction sh...

Страница 58: ...test can be performed with the device via the SebaKMT Cloud To save energy the loggers are not permanently ready to receive messages Only when this box is checked it is ensured that the device is avai...

Страница 59: ...or flow sensors the sign indicates the flow direction of the fluid Example 5 l s corresponds to 5 l s forward flow 5 l s corresponds to 5 l s reverse flow 3 A hysteresis and a debounce value can be de...

Страница 60: ...fied threshold 2 0 bar three times in succession 15 seconds 4 Select the days of the week alarming should be active on by marking the respective checkbox es The segment Alarm Trigger Exceptions offers...

Страница 61: ...g specify 4 00 as End time of monitoring tick the weekdays on which this monitoring alarm should apply here Monday to Friday Result If the flow value drops below 2 0 l s one or more times between 2 an...

Страница 62: ...tching input trigger an alarm The input voltage threshold is 2 5 V Any voltage value below 2 5 V is interpreted as 0 V while voltages higher than 2 5 V are interpreted as 5 V In order to ensure a reli...

Страница 63: ...up data communication in particular SebaKMT cannot provide any specific technical advice in this case For mobile communication a number of settings must be made Proceed as follows Step Description 1...

Страница 64: ...ctive checkbox Daily upload time Here you can specify at what time the measurement should be sent to the FTP server daily upload Select a time from the drop down list Each upload includes the newly re...

Страница 65: ...ng the configuration data the device starts recording data This is forwarded directly to the FTP server so that the progress of the measurement can be followed via the internet in real time If the dev...

Страница 66: ...s here to which summary or alarm messages should be sent E mail Destination Specify up to two e mail addresses here to which summary or alarm messages should be sent FTP Settings Enter the access data...

Страница 67: ...encrypted e mail transmission TLS SSL via an SMTPS e mail account If you want to use encryption check the Use TLS SSL box Ensure that the right port is specified in the e mail settings When using encr...

Страница 68: ...All addressees should receive a SMS or e mail containing the following information Type of device and identification number Date and time of the test Signal quality In the case of bad signal quality i...

Страница 69: ...provide information about this device or the location of the device etc 8 10 Sending the configuration data to the logger Finally the data must be transferred to the logger NOTE Keep in mind that the...

Страница 70: ...Programming the logger using the SebaDataView 3 software 70...

Страница 71: ...e live measuring values should be displayed graphically or numerically If you select Digital the measured value of each active channel is displayed numerically in a table If you select Graph the measu...

Страница 72: ...e 2 In the menu bar click on Read 3 Select Device from the appearing context menu Result The connection between the computer and the device is established and the data readout starts In a small window...

Страница 73: ...s not appear Instead the measurement data display opens immediately see page 74 The window contains a table in which all measurement data records saved for this device in the database of the software...

Страница 74: ...the list and click Show or double click the concerning line Result The measurement data is shown 9 4 2 Using the measurement data display The following figure shows the window for displaying measurem...

Страница 75: ...am you can restrict the time frame of the displayed area to a month a week a day or an hour By means of the User defined option you have the the chance to specify the start and end time of the span to...

Страница 76: ...tual diagram view right click in the bottom diagram and select Print from the appearing context menu If you want to cancel all active functions in the window and return to the initial diagram view rig...

Страница 77: ...you want to export in the directory tree 2 In the menu bar in the segment Measurement data click on Export Result An Explorer window opens 3 Browse to the desired target folder and save the data set...

Страница 78: ...Retrieving and evaluating data 78...

Страница 79: ...data exchange between device and computer If you wish more current information read the configuration of the device next step For that the following requirements must be met a radio interface must be...

Страница 80: ...evant device in the directory tree 2 In the menu bar in the segment Event list click on Show Result The event list window opens The events up to the time of the last data readout are shown If the list...

Страница 81: ...This means when starting the software you must have logged in with the password for User level 1 The User levels 2 and 3 are not authorized for password management To extend the SebaDataView 3 passwo...

Страница 82: ...an perform a reset with the device ie reset the logger to factory settings This deactivates password protection and makes the logger accessible again Note The reset deletes all measurement data stored...

Страница 83: ...proceed as follows Step Description 1 Download the file for the update from the download area of www sebakmt com onto your computer 2 Open the SebaDataView 3 software 3 Click the water drop symbol on...

Страница 84: ...7 Click OK to start the firmware update Result The update file is transferred to the device s and installed there A bar under the list on the right shows the progress of this process After the file is...

Страница 85: ...or contact us Contact information on our web site www sebakmt com Each replacement kit contains new O ring seals for the logger housing and the logger lid in addition to the batteries Replacing the se...

Страница 86: ...to the back and down Do not use force 5 Pull the battery out of the compartment 6 Insert the new battery Make sure that the wires of the battery are in the recess of the compartment and not be jammed...

Страница 87: ...ten 13 3 Reset of the battery counter After a battery replacement has taken place the internal battery counter of the device must be reset Otherwise the actual battery level cannot be recognized by th...

Страница 88: ...Changing the battery 88...

Страница 89: ...triche vecchie z me nor da ka iztr d jumu uz kura t atrodas nedr kst izmest kop ar parastiem m jsaimniec bas atkritumiem T k tas ir izstr d jums ko cits citam p rdod un lieto tikai uz mumi tad to nedr...

Отзывы: