SetMeasIntervalSer#=x Interval, in seconds, between each command string. Set by SetMeasStringSer#= to serial sensor
during 25plus data collection. If SetMeasStringSer#=0, the 25plus sends a command string set by
SetMeasStringSer#= as soon as it receives a response to the last command for data.
SetStartStringSer#=x
Command string, 0–32 characters, that the 25plus sends to the serial sensor for it to collect data
autonomously during data collection. If empty, push
Enter
. No command will be sent to the serial
sensor. This is appropriate for a serial sensor that must be commanded to take each sample (see
SetMeasStringSer#= above) or that automatically collects a sample and sends data when power is
supplied. Default: Start
SetExecCharSer#=x
Decimal value, 0–255, of the command execution character that the 25plus adds to the end of the
command string. Refer to SetMeasStringSer#= and SetStartStringSer#=.
254 is carriage return line feed ('\r\n')
255 is no execution character.
Default: 254
SetTermCharSer#=x=
Decimal value, 0–255, of termination character that the serial sensor sends to the 25plus at the end
of its response to a command. When the 25plus receives a termination character, it ignores any
additional characters from the sensor.
254 is carriage return line feed ('\r\n')
255 is no termination character.
If the 25plus does not receive a termination character, it no longer waits after SetFailOutSer#=.
Default: 254
SetSuppressSer#=x
x=Y: Default. The termination character, SetTermCharSer#= and prompt, SetPromptSer#= sent
from the serial sensor to the 25plus do not show in the data file.
x=N: The termination character and prompt sent from the serial sensor to the 25plus show in the
data file.
SetTimeoutSer#=
Maximum time, in seconds, that the 25plus will wait for prompt (SetPromptSer#=) from a serial
sensor that it has completed a command that was sent. Range: 1–600 seconds. Default: 60
SetFailoutSer#=x
Maximum time, in seconds, that the 25plus will wait for data from a serial sensor. Range:
1–600 seconds. Default: 60
ToSer#=x
Command string, 1–64 characters, to send to serial sensor. This can be any command used by the
serial sensor.
Notes:
•
The 25plus cannot transmit real-time data from a serial sensor.
•
The 25plus configuration file (.xmlcon) must agree with the SetInlineSer#= value
whether data is serial sensor data is stored in-line with other data or in a .txt file to
process data in the SBE Data Processing software. Look at and edit the configuration
file in the software.
8.6 Real-time output setup
The 25plus always operates at 16Hz, but the output rate of real-time data changes. The
rate at which data can be transmitted is a result of the baud rate and quantity of auxiliary
voltage sensor data transmitted in real-time. The maximum rate of output is 8 Hz (8 scans
per second).
Reference: command descriptions
48
Содержание SBE 25plus
Страница 2: ......
Страница 6: ...SBE 25plus quick start guide 4 ...
Страница 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Страница 14: ...Figure 8 SBE 55 Overview 12 ...
Страница 22: ...Set up sensor and verify operation 20 ...
Страница 60: ...Serial sensor integration 58 ...
Страница 64: ...General information 62 ...
Страница 65: ......