e.
Align the Tygon tubing on the joint between the C duct and T duct.
11.
Install the conductivity and temperature sensors and the mounting bracket on the
main housing.
12.
Connect the quick-release plug on the conductivity cell.
13.
Connect the cables and lock collars.
7.6 Replace batteries
W A R N I N G
Explosion hazard. If the batteries are not installed correctly, explosive gases can be
released. Make sure that the batteries are of the same approved chemical type and are
inserted in the correct orientation.
W A R N I N G
If the user thinks that a sensor has water in the pressure housing: Disconnect the sensor from any
power supply. Put on safety glasses and make sure that the sensor is pointed away from the body
and other people. In a well ventilated area, use the purge port (if the sensor is so equipped), or
very SLOWLY loosen the bulkhead connector to let the pressure release.
W A R N I N G
If the user thinks that the alkaline batteries have leaks, pressure may have built up inside of the
pressure housing. Follow ESD protocols to release internal pressure. Put on safety glasses and
protective gloves and make sure that the sensor is pointed away from the body and other people.
In a well ventilated very SLOWLY loosen the bulkhead connector to release the pressure. Keep
away from heat, sparks, flame, and other sources of ignition. Do not smoke.
W A R N I N G
This product can expose the user to chemicals with silica, crystalline (airborne particles of
respirable size), which is known to the State of California to cause cancer and birth defects or
other reproductive harm. For more information, go to
C A U T I O N
The pressure housing contains Electrostatic Discharge (ESD) sensitive parts and assemblies that
are susceptible to damage from ESD. Follow ESD protocols.
N O T I C E
To SLOWLY remove the bulkhead connector: loosen one bulkhead connector by 1 turn to open
the O-ring seal under the connector. If there is internal pressure, water may leak out or you may
hear a hiss. Let the pressure bleed off slowly. Then it will be safe to remove the end flange.
Maintenance
34
Содержание SBE 25plus
Страница 2: ......
Страница 6: ...SBE 25plus quick start guide 4 ...
Страница 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Страница 14: ...Figure 8 SBE 55 Overview 12 ...
Страница 22: ...Set up sensor and verify operation 20 ...
Страница 60: ...Serial sensor integration 58 ...
Страница 64: ...General information 62 ...
Страница 65: ......