Section 7 Maintenance
7.1 Corrosion precautions
N O T I C E
Blue Moly
™
and Dow Corning
®
4 lubricants are electrically conductive. Keep away from electrical
components.
Flush the 25plus with fresh water after each use and before it is stored.
The stainless steel screws that connect the plunger switch to the end flange are
lubricated with a thick layer of Blue Moly
™
, and all other stainless steel screws that are
exposed to salt water are lubricated with a thick layer of Dow Corning
®
4. Remove and
lubricate all of these screws after each deployment.
The 25plus has a large zinc anode attached to each end flange. Make sure to examine
these at regular intervals to verify that they are securely attached and in good condition.
Replace when corrosion has reduced either one to less than 50%.
Do not attach metal objects directly to the pressure housing.
7.2 Plumbing
Clean the Tygon tubing at regular intervals to prevent clogged plumbing or air release
valve. To clean the air release valve:
1.
Remove the Tygon tubing above the air release valve and use needle-nosed pliers to
push a 0.4 mm diameter wire (or #26 AWG wire) through the valve hole.
2.
Blow air through the air release valve to make sure it is open.
3.
If necessary, replace the Tygon tubing above the air release valve.
7.3 Clean pressure sensor
N O T I C E
Do not put a brush or any object in the pressure port. It may damage or break the pressure
sensor.
31
Содержание SBE 25plus
Страница 2: ......
Страница 6: ...SBE 25plus quick start guide 4 ...
Страница 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Страница 14: ...Figure 8 SBE 55 Overview 12 ...
Страница 22: ...Set up sensor and verify operation 20 ...
Страница 60: ...Serial sensor integration 58 ...
Страница 64: ...General information 62 ...
Страница 65: ......