Approximately one time each year, examine the pressure port for particles and debris.
Flush the pressure port with warm de-ionized water to clean.
7.4 Conductivity cell
N O T I C E
Do not put a brush or any object inside the plumbing to clean it. The conductivity cells can be
damaged, which changes the calibration.
Do not store the sensor with water in the plumbing. The conductivity cells can break and the
oxygen sensor can be damaged if they freeze.
Note:
The manufacturer ships the SBE 4C conductivity cell with no water in the cell so that it
does not freeze.
Use the quick-disconnect to disconnect the conductivity cell, then use the manufacturer-
supplied conductivity cell kit to clean the cell.
7.5 Disassemble and reassemble TC duct
N O T I C E
Use slow, deliberate movements to disassemble or assemble the TC duct to prevent damage to
the temperature or conductivity sensor. Do this procedure in a laboratory, not on a ship.
Prepare the temperature and conductivity duct to send back to the manufacturer for
calibration, or to clean. Disconnect the C duct from the T duct and remove both from the
conductivity cell. Send the temperature sensor with the guard attached to prevent
damage to the thermistor.
1.
Put the system cage on its side.
2.
If necessary, disconnect the bulkhead connectors from the temperature and
conductivity sensors.
Maintenance
32
Содержание SBE 25plus
Страница 2: ......
Страница 6: ...SBE 25plus quick start guide 4 ...
Страница 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Страница 14: ...Figure 8 SBE 55 Overview 12 ...
Страница 22: ...Set up sensor and verify operation 20 ...
Страница 60: ...Serial sensor integration 58 ...
Страница 64: ...General information 62 ...
Страница 65: ......