Section 3 Overview
The battery-powered 25plus measures temperature, conductivity, and pressure and
supports up to 8 voltage-output sensors and 2 RS232 sensors. Components include:
•
SBE 3F temperature sensor
•
SBE 4C conductivity sensor
•
SBE 5 pump
•
Eight 16-bit A/D channels for auxiliary voltage output sensors for dissolved oxygen,
pH, fluorescence, PAR, light transmission, turbidity, and other measurements
•
Two channels for auxiliary serial output sensors
•
Stainless steel cage
•
Optional: strain-gauge pressure sensor in eight depth ranges.
The 25plus has a 16 Hz scan rate for high spatial resolution and can operate for up to
55 hours (without auxiliary sensors). It records and stores data that is transmitted through
RS232 or USB to a PC. Data can also be transmitted in real-time from 0.5 to 8 Hz.
Data storage
Example 1
2GB memory, no auxiliary sensors. T, C, and P = 10 bytes/sample
storage: 2,000,000,000 ÷ 10 = 200,000,000 samples
Example 2
2GB memory, 4 auxiliary sensors. T, C, and P = 10 bytes/sample
External voltages = 2 bytes per sample × 4 sensors = 8 bytes per sample
storage: 2,000,000,000 ÷ (10 + 8) ~ 111,111,111 samples
The 25plus can be integrated with an SBE 32 Carousel Water Sampler or an SBE
55 ECO Water Sampler. It can also be used with an SBE 36 Deck Unit and Power/Data
Interface Module (PDIM) or SBE 33 Deck Unit and water sampler for real-time operation
with cable lengths up to 10,000 m.
3.1 Integration with other equipment for real-time data collection
The 25plus can be used with any of the equipment listed below:
•
SBE 32 Carousel Water Sampler (12, 24, or 36 bottles) and SBE 33 Carousel Deck
Unit
•
SBE 36 CTD Deck Unit and PDIM
•
SBE 55 ECO Water Sampler (3 or 6 bottles) and SBE 33 Carousel Deck Unit.
All of these supply a minimum of 15 VDC to the 25plus, and supply sufficient power for
auxiliary sensors. The bottles on the SBE 32 and SBE 55 and be closed at any depth and
will not have an effect on data collection.
9
Содержание SBE 25plus
Страница 2: ......
Страница 6: ...SBE 25plus quick start guide 4 ...
Страница 12: ...Figure 4 SBE 32 Figure 5 SBE 36 Overview 10 ...
Страница 14: ...Figure 8 SBE 55 Overview 12 ...
Страница 22: ...Set up sensor and verify operation 20 ...
Страница 60: ...Serial sensor integration 58 ...
Страница 64: ...General information 62 ...
Страница 65: ......