NOVA
WAVE
FA
Place the
NOVA
WAVE
SA in the position where you plan to use the system allowing at least 28 inches (72 cm) on each side for the
Transporter.
Use a straight edge to ensure the top of the Transporter is level with the bottom of the tunnel. This is accomplished by placing the straight
edge in the tunnel parallel to the back wall of the tunnel. Using the leveling feet, adjust the system to the correct height by turning the feet.
Clockwise lowers the table top and counter clockwise lifts the Transporter top. Adjust the back feet first to the point where the straight edge
touches the top of the Transporter top along the entire length of the straight edge. Do this on both the right and left hand sides. Position the
right and left hand sides into place by ensuring the tongues of the Transporter tops are centered within the door openings. Use the straight
edge along the back wall of the tunnel and the back rail of the left hand Transporter module to ensure that the Transporter it is aligned along
the length of travel.
Connect the bridge between the left and right side Transporter modules and ensure that the alignment pins are pushed together snuggly. Raise
or lower the feet on the front of the Transporter to level the right and left Transporter modules.
Install the cooling and venting module on the right hand side of the
NOVA
WAVE
. Install the cooling duct by screwing it in the threaded hole
on the right side of the SA. Connect the flexible duct to a fume hood.
NOTE
It is important to ensure that the exhaust hose is vented into a fume hood or to the outside.
Install the left hand cover unit using the hole and guide pins for alignment.
Precautions To Avoid Possible Exposure To Excessive Microwave Energy
(a) Do not attempt to operate this instrument with the door open since open-door operation can result in harmful exposure to microwave
energy.
It is important not to defeat or tamper with the safety interlocks.
(b) Do not place any object between the instrument front face and the door or allow soil or cleaner residue to accumulate on sealing
surfaces.
(c) Do not operate the instrument if it is damaged. It is particularly important that the tunnel door close properly and that there is no damage
to the:
(1 ) Door (bent), (2 ) door seals and sealing surfaces.
(d ) The instrument should not be adjusted or repaired by anyone except properly qualified service personnel.