Version 0.15.1
NOVA
WAVE
- A New Category of Automated Microwave Digestion
|
27
With a
NOVA
WAVE
SA unit the
cooling of the rack may take place
in the tunnel either By time or By
temperature.
Cooling is to determine where the
first stage of cooling takes place. If
By time is selected, it will use the
internal exhaust fan to cool for
the time period entered. If By
temperature is selected, the
system will monitor and wait until
all samples are cooled to this set
temperature.
The system cannot cool below ambient temperature.
If zero time is selected the racks will move to the out position after the last step in the heating profile is complete.
OK saves the changes made and returns to the previous page.
Cancel returns to the previous page without saving the changes made.
A Name for a method must be populated. Remarks are used for additional comments. These comments will be printed on the
reports page.
NOTE