Safety messages may occur during system operation due to a variety of conditions as included below. For further assistance, consult your lab
manager or
SCP SCIENCE
Service Department or your local distributor.
List of Safety Messages:
1.
Tube Number # did not reach target temperature
a. Safety pressure cap not sealed correctly
b. Sample tube vented
c. Time to temperature plus time at temperature has been exceeded without reaching set point temperature
Solution: increase time to temperature
2.
Temperature is rising to slowly in the tube number #
a. Sample vented
b. Pressure cap not sealed properly
c. Applicators not closed properly
d. Magnetron error
e. Thermopile error
3.
Microwave cable number # overheats
a. Tube has boiled dry
b. Insufficient acid in vessel
c. Applicator failure
4.
Temperature in tube number # is rising to quickly
a. Exothermic reaction
5.
Alarm from High Voltage PCB number #
a. PCB over heating
b. Fans stopped running
c. Air vents plugged
d. Cooling fans not running
Safety Messages
CAUTION
If the reason for the safety message is unclear contact
SCP SCIENCE
Service Department or your
local distributor.
S
afety
M
eSSageS
froM
the
P
ower
M
anageMent
B
oard
1.
Power Management has reset
2.
Thermal Switch has failed
3.
24V supply has failed
4.
Keep Alive signal from FPGA has been lost
5.
5V supply has failed
6.
MicroBlaze is updating
7.
GUI is shutting down
If any of the safety messages are displayed in RED on the touch screen, please call your local distributor or contact
SCP SCIENCE
’s Service
Department. Contact information is located at the back of this manual.