Version 0.15.1
44
|
9
Troubleshooting
No. Error Message
Description
1
COM.ERROR
Unrecognized packet.
2
ERROR! ID=Stalled DESC= Section 1 (2,3,4) is stalled. FA
Rack is stalled at B-F section.
3
ERROR!!! ID= Timed Out. DESC= Section Operation has timed out. FA
Operation B-F timed out. (R-L, F-B or L-R).
4
ERROR COM Timed out! FA
No response from MB.
5
ERROR No OK received from MB. FA
OK not received from MB.
6
Process terminated with exit code #. Press OK to restart.
GUI crashed code number #. Press OK to resume.
7
DIGESTION.RUN TUBE 01 E_CODE=02
Never reached target temperature.
8
DIGESTION.RUN TUBE 01 E_CODE=03
Temperature rising too slowly in tube number #.
9
DIGESTION.RUN TUBE 01 E_CODE=04
Tube number # overheats.
10
MB_SYNTH
System initialize.
11
ERROR Digestion Operation has timed out!
Digestion operation has timed out on tube number #.
12
ERROR DIGESTION_ALARM_ERROR Magnetron is alarmed.
Magnetron number # has an alarm.
13
DIGESTION_OVERHEAT_ERROR Magnetron overheated
Magnetron number # overheats.
14
ACTUATOR_STALLED_ERROR
Actuator is stalled.
15
ACTUATOR_OPERATION_TIMEDOUT_ERROR
Actuator operation has timed out.
16
CAVITY_STALLED_ERROR
Tunnel motor is stalled.
17
CAVITY_UNKNOWN_RACK
Please remove unknown rack from tunnel.
18
CAVITY_OPERATION_TIMEDOUT_ERROR
Tunnel operation has timed out.
19
CAVITY_OBSTRUCTED_ERROR
Tunnel is obstructed.
20
CAVITY_NOT_INITAILIZED_ERROR
Tunnel motors need initialization.
21
CAVITY_NOT_NOW_ERROR
This command cannot be performed.
22
CAVITY_ORITENTATION_FAILED_ERROR
Unknown rack orientation.
Messages may occur during system operation due to operator error or component failure. For further assistance, contact
SCP SCIENCE
Service Department or your local distributor.
List of Error Messages:
23
CAVITY_INTERNAL_ERROR
An error occurred in tunnel.
24
RACK ID FAILED
25
RACK IN REVERSE POSITION
26
PMB
Thermal switch has failed.
Содержание NOVAWAVE
Страница 1: ...NOVAWAVE AnewcategoryofAutomatedMicrowaveDigestion User Manual ...
Страница 45: ...Version 0 15 1 45 NOTES ...