Version 0.15.1
NOVA
WAVE
- A New Category of Automated Microwave Digestion
|
17
To program a digestion by rack
position, a cycle with racks must
be created first. Highlight the rack
number with the different samples to
digest. Select the Sample List button.
On this page, to select a position
to program, select all of same type
of samples by clicking on the rack
position (
Pos
) button. Select a method
from the Method pull down tab; select
a vessel type from the Vessel Type
pull down. Click on the Enter Value
tab to populate table. If a second
method is required select the position
to program in the same manner as
above. Method and vessel type may
vary by position. If the vessel position
is not populated the
NOVA
WAVE
will consider that there is no vessel in
that position.
After a method is selected, a vessel
type needs to be selected for the
system to operate correctly. Select
one of the pre-programmed vessel
types from the Vessel Type pull down
tab. If the Cycle is highlighted, the
vessel type will be assigned to all of
the racks for this cycle.
Quartz vessels are either 50 ml or
75 ml vessels.
Fluoropolymer vessels are either 50
ml or 75 ml vessels.
Other is reserved for future use.
Notube is selected if the position in
the rack is intentionally left blank.
Vessel selection is used to set the IR
temperature sensor parameters.