![scigiene HM-14560 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/scigiene/hm-14560/hm-14560_user-manual_1217606016.webp)
Page 16
5.
Care and maintenance
Regular cleaning and maintenance will ensure that your device remains in perfect condition as long as
possible.
ATTENTION
Sensitive sensor plate
The white sensor surface consists of a thin ceramic plate. This can be destroyed
by mechanical pressure!
►
When installing the sensor, remove the provided transport protection as
late as possible!
5.1 Adjustment of the device
Every two weeks an adjustment must be run (adjust). This procedure guarantees a long term stability of
the moisture measuring.
Requirement:
The device must have a temperature between 20.0 °C and 26.0 °C and must be on for at
least 30 minutes.
1.
Remove the paper or cardboard under the sensor surface.
»
Make sure there is nothing under the sensor sur- face.
2.
Press the button with the red LED until the LED lights up
constantly (image 8).
»
When you release the button, the adjustment begins!
»
During adjustment, the red LED flashes every second.
3.
Once the adjustment is completed, the red LED goes out for
a moment and then flashes again every 5 seconds.
5.2
Reset
•
Press the button with the green LED (image 9).
»
The button with the green LED lights up for one second and the
button with the red LED flashes rapidly for this period.
•
The device will now restart.