3.7
Pin assignment Modbus TCP (option)
A 4-pole M12 connector with D-coding (M12-D) is required here.
•
A CAT6 or higher type cable is recommended.
•
Every cable type must be equipped with a shielded M12 round plug D-coded on the sensor side.
»
The cable shield must be earthed in this installation setup!
3.8
Laying the power supply or transmission line
•
The cable cannot be laid in the range of interference fields.
•
The transmitter cannot be used near electromagnetic interference fields.
•
The cable should not be strongly bent.
•
Permissible cross-sections for installation have to be considered.
•
The cable length must be kept as short as possible.
»
Should an extension be necessary, the cross-section of the extension may not be under 0.25 mm².
»
The potential equalization GND ( shielding ) must be connected.
Pin No
Function
1
TX+
2
RX+
3
TX-
4
RX-
Housing
Shielding