3.6
Pin assignment power supply and analogue output (option)
Cable color
Pin No
Function
Brown
1
Power supply V- (0 VDC)
White
2
Power supply V+ (24 VDC)
Blue
3
Power output moisture 4 - 20 mA (optional)
Black
4
Ground power output & temperature (optional)
Grey
5
Power output temperature 4 - 20 mA
(optional)
Purple
Housing
Potential equalization GND
(
shielding
)
ATTENTION
Measuring signal interference caused by missing potential equalization
A missing potential equalization can transfer interference signals to the measuring signal.
►
The potential equalization GND (shielding) must be connected.
Electronic damage caused by incorrect cable connection
Incorrect assignments can cause serious electronic damages.
►
Connect all cables correctly.
Electrical damages caused by extension problems
Extension problems can cause electronic damages.
►
The cross-section of the extension may not be under 0.25 mm2.