Maintenance 

The SliceMate should be regularly maintained to ensure reliable operation; the 
frequency of maintenance is dependant on the type of compounds used. It is 
recommended that all replaceable components listed below be simultaneously 
exchanged on a regular basis with exception of the brain bath module. 

 

AgCI Electrode Assembly 

 

Heat Exchanger Tube 

 

Pinch Valve Tube 

 

Feed through barbed connectors 

 

Brain bath mesh 

Replacing the Brain Bath Mesh 

Firstly, carefully ease the mesh ring from the brain using a scalpel and remove the 
ring and the mesh. Cut an oversize piece of mesh and lay it over the bath area as 
shown.  

 

 

Now press the ring back into the bath; this action stretches the mesh to create a 
stable interface for the slice to sit upon. 

 

 
 
 
 

Содержание SliceMate

Страница 1: ...1 SliceMate Manual...

Страница 2: ...e housing integrates the tissue bath heat exchanger Peltier device and fluidic control valve within a dedicated Acetyl enclosure The compact and precision design ensures system variables are minimised...

Страница 3: ...he Stainless Steel ground plane of the SliceMate The heat exchanger spiral is held in position by a stainless steel spring clip this ensures the heat exchanger has ideal contact with the Peltier devic...

Страница 4: ...a low resistance earth bond via the connector shells Reference Electrode Ground The reference electrode ground is located in between the primary well and the suction port in the brain bath The head st...

Страница 5: ...livery The system can be used with either a gravity fed or pumped system The system is designed to operate over a flow rate range of 1 to 10 ml minute The fluid should be delivered to the SliceMate us...

Страница 6: ...the vacuum The bath solution height is maintained and adjusted using the vacuum wicking method The SliceMate bath has a pair of oval suction vents which are connected to the outlet port at the rear of...

Страница 7: ...done so that when the flow is turned back on there will be a minimal risk of a hot surge of fluid entering the bath and the desired temperature is quickly regained The system initialises on power up...

Страница 8: ...low and allow the target temperature to stabilise Continue with the experiment End Experiment Turn the thermal control off by selecting the station number on the console handset and pressing stop Note...

Страница 9: ...ith exception of the brain bath module AgCI Electrode Assembly Heat Exchanger Tube Pinch Valve Tube Feed through barbed connectors Brain bath mesh Replacing the Brain Bath Mesh Firstly carefully ease...

Страница 10: ...oved from the SliceMate base Unscrew the two black Acetyl knobs in an anticlockwise direction and lift the SliceMate clear from the table Replacing the heat exchanger tubing 1 Firstly obtain a piece o...

Страница 11: ...led free 6 Now install the new tubing take the angled cut end have this extending from the edge of the spiral by 6 5 to 7 5mm to allow attaching the silicon valve tubing later on Being careful not to...

Страница 12: ...es lifting the brain slice during operation Remove the tubing from the inlet port to the heat exchanger by first pulling it out of the pinch valve and then pulling away from the heat exchanger and fin...

Страница 13: ...ng 5 Now avoiding kinking the heat exchanger tubing form a comfortable transition from the heat exchanger into the valve jaw 6 Using a spatula or similar tool press the silicon tubing into the valve j...

Страница 14: ...venient replacement with minimal tooling The silver chloride pellet is sleeved within silicon tubing this acts as both a seal between the pellet and brain bath and as a pull sleeve enabling easy insta...

Страница 15: ...from the top side of the bath and then gently pulled from the underside until the surface of the pellet is level with the surface of the bath If the user prefers to make the assembly themselves the fo...

Страница 16: ...or wire polythene tube assembly through a new piece of silicon tube of 35mm length 6 Solder the connector wire to AgCI Electrode ensure smooth streamline joint 7 Ensure polythene sleeve slides freely...

Страница 17: ...Heat Exchanger Lid Retainer Screws Tube Cutter Silicon Tubes Steel Rule Measuring Tubing Length s Soldering Station Ref Electrode Off line Assembly Solder 5 core Ref Electrode Off line Assembly Plast...

Страница 18: ...um cured silicone return heat exchanger to Outlet Port 1 8 ID platinum cured silicone Bath Earth Ag Cl pellet E201 33 0003 Harvard Bath Earth tube seal Silicone tubing 125 OD 125 ID Thermistor ALA Bea...