2.
Turn the SliceMate back over and unlock the brain bath from the bayonet
mechanism and lift the bath free.
If replacement silver chloride electrode assemblies are purchased from Scientifica
then the old electrode can simply be removed and discarded. The new electrode
is then fed from the top side of the bath and then gently pulled from the underside
until the surface of the pellet is level with the surface of the bath.
If the user prefers to make the assembly themselves the following procedure must
be undertaken:
1.
Disconnect and pull old electrode from base of brain bath
2.
Remove the silicon tubing and discard.
3.
Slide the polythene tubing back from the smooth solder connection
4.
De-solder the connecting wire from the earth pellet and discard the pellet