14| 4.0 Operation Guide
M+ - Pressing the M+ button stores the current position of the MMTP in
memory. Moving the MMTP and pressing M+ again will store another
position. Up to 25 memory positions can be stored.
MR - Pressing the MR button will recall the first stored memory position.
If more than one position is stored subsequent presses of the MR button
will recall each position stored in the same sequence they were saved.
Note: To clear stored memory positions, simultaneously press
the M+ and the MR buttons. This will remove all stored positions
from memory.
H in & H out – These buttons allow two positions to be stored in
memory, the Home In position and the Home Out position. The Home
Out position must be set first before Home In can be used.
To set Home Out move the MMTP to the desired position then press H
out for 3 seconds to store the current position. Subsequent pressing of
Home Out will cause the MMTP to move out to the stored Home Out
position until an alternative Home Out position is stored.
The Home In position can be set in two modes:
Mode 1 – This is the default mode. The Home In position can only be
set once a Home Out position is set. From the Home Out
position, pressing the Home In button will cause the MMTP to return its
original position prior to Home Out, minus a safety minus a safety offset.
This value can be configured using the LinLab control software.
Mode 2 - This allows the user to set a specific position as the Home In
point. Move the MMTP to the desired Home In position then hold down
the Home In button for 3 seconds. The current position is stored to
memory as the Home In position. No Home in offsets will be applied in
Mode 2.
Note: If the Home Out and Home In buttons are held down
simultaneously, the Home Out and Home In positions will be
cleared and the Home In will be set back to Mode 1.