5. Instructions for Use
5.2 Cycle Description Chart
Page 7
* This cycle is not a wash cycle and is not suitable for processing instruments prior to
reuse. Always run a wash cycle following the rinse and hold cycle.
** Cycle times depend on the temperature of incoming water.
*** Cold prewash water temperatures as stated is for normal operations allowing units to
cool down between cycles for a minimum of 30 minutes.
P1 – rinse and
P2 – regular
P3 – Heavy duty
Hold* cycle
cycle
cycle
description
cold Prewash***
<30ºC / 86ºF
<30ºC / 86ºF
<30ºC / 86ºF
Wash
N/A
50ºC / 122ºF
50ºC / 122ºF
5 minutes
9 minutes
rinse
60ºC / 140ºF
60ºC/ 140ºF
60ºC / 140ºF
dry
N/A
1-15 min.
1-15 min.
(10 min. default)
(10 min. default)
Total Time
**
8 minutes
19 minutes
23 minutes
without drying
Water
6 L / 1.59 gallons
17L / 4.5 gallons
17L / 4.5 gallons
consumption
Use to prevent soil
from drying on
instruments when
they will not be
washed within
one hour.
Use for moderately
soiled loose
instruments.
Use for heavily
soiled instruments
and cassettes