5.1 Instrument Reprocessing Recommendations
The HYDR
IM
C51w is designed to clean general dental and medical instruments.
SciCan recommends terminal sterilization after processing in the HYDR
IM
C51w.
When used correctly, the HYDR
IM
C51w is effective for the removal of debris on
instruments that may result in the insulation of microorganisms during terminal
sterilization. Prior to placing instruments in the HYDR
IM
, consult the instrument
manufacturer’s reprocessing instructions.
Examples of medical and dental instruments suitable for cleaning in the HYDR
IM
C51w include:
• Stainless steel scalers, spatulas and hemostats.
• Stainless steel retractors, needle holders, skin hooks, tissue forceps, scissors,
curettes, and other stainless steel instruments free of deep cavities.
For best results, instruments processed in the HYDR
IM
C51w should
have the following properties:
• Heat resistance to a temperature of up to 70°C / 158°F
• Corrosion resistance in the presence of heat and alkalinity.
Dental handpieces, microkeratomes, phaco tubing, fiberoptics and electrical equipment
should not be processed in the HYDR
IM
.
HIP
TM
(HYDR
IM
Cleaning Solution with Instrument Protection) has been carefully
formulated to provide outstanding material compatibility with a wide range of metals and
coatings. However, some manufacturers’ instruments may not be compatible.
All hinged instruments should be placed in open baskets in Heavy Duty cycle.
When processing small or light objects in the HYDR
IM
use the basket with lid
(SciCan part number 01-108294).
5. Instructions for Use
Page 6