5. Instructions for Use
The cycle will start by draining the unit for 20 seconds. The bar
on the screen indicates the progression of the cycle. The stage
within the cycle is shown beneath the bar (eg. Filling, Rinsing).
The estimated time remaining (in minutes) is always displayed.
Please note that the time will increase or decrease depending on
the projected cycle completion.
When the cycle is complete, the program name and “Cycle
Complete” is displayed, and the unit beeps three times.
Press to open the door. Use gloves to unload the
instruments. After all cycles, exercise caution as the
instruments, trolley, baskets and cassettes may be hot. Inspect
the instruments for debris prior to packaging / sterilizing.
5.5 Interrupting / Aborting a Cycle
The door of the HYDR
IM
is locked during the cycle. If the
STOP
button is pressed, the cycle is aborted and cannot be re-started.
The message “Draining. Please wait” will appear. The unit will
drain. Then the message “Instruments not processed. Cycle
Aborted” will appear. Press OK. Open the door by pressing .
Use gloves if handling the instruments. Exercise caution as
instruments, trolley, baskets and cassettes may be hot.
The cycle must be repeated to ensure that instruments are
properly processed. Note that if the drying phase of the cycle
is interrupted, the critical parts of the cycle are complete
and the message displaying the program name and “Cycle
Complete. Drying interrupted” will appear.
Page 10
P2 Wash
FILLING
19 min
P2 Wash
Cycle Complete
P2 Wash
Instruments
not processed
Cycle Aborted