Bravo G4 17 / Bravo G4 22 / Bravo G4 28
EN
83
CODE
POSSIBLE CAUSE
SUGGESTED SOLUTION
MIN. water level sensor failure.
Contact Technical Service
(see Appendix).
E031
Water level in the drain tank over the MAX. level.
RESET according to the instructions and empty the tank.
Completely drain the drain tank.
MAX. water level sensor failure.
Contact Technical Service
(see Appendix).
Problem in the hydraulic circuit.
E042
Warning that the maximum water level in the tank has
been reached (manual filling)
Interrupt the filling operation to prevent water spillage.
E050
Reminder to run the Vacuum Test cycle
Run Vacuum Test as planned
E051
Reminder to run the Helix Test cycle
Run Helix Test cycle as planned
E052
Reminder to run the combined Helix Test
cycle
Run Helix Test combined cycle as planned
E060
The autoclave cannot connect to Lan network
Make sure that configuration parameters of the Lan network are correct. Check
that the Lan network chosen for the connection is working properly.
Contact Technical Service (see Appendix).
E061
The autoclave cannot connect to WiFi network
Make sure that configuration parameters of the WiFi network are correct. Check
that the router managing the WiFi network is on and that the Wi-Fi network chosen
for the connection is working properly.
Contact Technical Service (see Appendix).
E126
Cloud firmware is being updated
Wait for the message to disappear and restart the machine
E141
The cloud firmware version is not the correct one with
respect to the firmware process. There may be
malfunctions in the connection with WiFi / ethernet or
cloud
Contact Technical Service
E900
Air seepage through the gasket.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Problem in the hydraulic circuit.
Contact Technical Service
(see Appendix).
E901
Excessive humidity in the sterilisation chamber.
RESET according to the instructions.
Thoroughly dry the inside of the chamber and restart the program.
Air seepage through the gasket.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Problem in the hydraulic circuit.
Contact Technical Service
(see Appendix).
E902
Excessive humidity in the sterilisation chamber.
RESET according to the instructions.
Thoroughly dry the inside of the chamber and restart the program.
Air seepage through the gasket.
RESET according to the instructions.
Thoroughly clean the gasket with a clean cotton cloth moistened with water.
Restart the program.
Vacuum pump failure.
Contact Technical Service
(see Appendix).
Problem in the hydraulic circuit.
E998
Service maintenance in progress.
Service maintenance in progress. If you were not informed, contact
IMMEDIATELY the manager of the network to which the steriliser is connected.
Contact Technical Service (see Appendix).
E999
Manual interruption of the sterilisation or test cycle.
RESET according to the instructions.