14. APPENDIX - MAINTENANCE
In addition to correct use, the user needs to perform regular maintenance in order to guarantee safe, efficient operation
over the device’s entire life.
Always use personal protective equipment.
For better quality of maintenance, supplement routine checks with regular periodic check-ups that can be performed by Technical Service Department
(see Appendix)
.
It is also fundamental to perform a
periodic steriliser validation,
i.e. a check of process thermo-dynamic parameters and their comparison with the
reference values detected by duly calibrated tools. Refer to 'Steriliser periodic validation' in the next part of the Appendix.
The regular maintenance described below consists in easy manual operations and preventative interventions involving simple tools.
In the event of replacement of components or parts of the device, request and/or use original spare parts only.
14.1. REGULAR MAINTENANCE PROGRAMME
The table summarises the maintenance interventions required to maintain the steriliser in good working order.
In case of
heavy use
, we recommend to
shorten
maintenance intervals:
DAILY
Clean the door gasket and the internal part of the door (14.3.1)
Clean the chamber drain filter (14.3.4)
WEEKLY
Clean external surfaces (14.3.3)
Clean the sterilisation chamber and its accessories (14.3.2)
Clean the anti-dust filter (14.3.6)
PERIODICALLY
Warnings related to Scheduled Maintenance are displayed with the following frequency (14.2):
WARNING MESSAGE
FREQUENCY
CHAMBER FILTER CLEANING (14.3.4)
250 CYCLES
DOOR LOCK LUBRICATION (14.3.5)
250 CYCLES
ANTI-DUST FILTER CLEANING (14.3.6)
500 CYCLES
BACTERIOLOGICAL FILTER REPLACEMENT (14.3.7)
500 CYCLES
DOOR GASKET REPLACEMENT (14.3.8)
1000 CYCLES *
GENERAL SERVICE
3000 CYCLES
YEARLY
Validate steriliser (14.4)
Replace the door gasket (14.3.8) *
* Replacement every 1000 cycles or 1 year, whichever is first.
A regular maintenance is essential to achieve the best performance of the device.
Periodically, a message will be displayed requesting that the above maintenance operations are performed.
For further information or in case of doubt, contact the technical service: if they have performed regular maintenance on the device, the technician
might have already carried out some of these operations (e.g. >Replacement of the bacteriological filter or of the gasket).