Bravo G4 17 / Bravo G4 22 / Bravo G4 28
EN
77
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET MODE
E141
The cloud firmware version is not the correct one with respect to the
firmware process. There may be malfunctions in the connection with
WiFi / ethernet or cloud
INCORRECT FW VERSION CLOUD
PLEASE UPDATE THIS FW
1
E900
Vacuum test failed
(during TEST PHASE)
TEST FAILED
SECOND STEP
CONTACT TECH.SERVICE
2
E901
Vacuum test failed
(during STAND-BY PHASE)
TEST FAILED
FIRST STEP
CONTACT TECH.SERVICE
2
E902
Vacuum test failed
(vacuum pulse time-out exceeded)
TEST FAILED
VACUUM NOT ACHIEVED
CONTACT TECH.SERVICE
2
E998
Remote maintenance activity in progress
REMOTE SERVICE
ACTIVE
1
E999
Manually interrupting the cycle
MANUAL INTERRUPTION
2
1 = OK (warning)
2 = OK + door unl RESET (if in cycle)