Bravo G4 17 / Bravo G4 22 / Bravo G4 28
EN
79
CODE
ALARM DESCRIPTION
MESSAGE ON THE DISPLAY
RESET MODE
A135
Solenoid valve 5 failed
SOLENOID VALVE 5 ERROR
CONTACT TECH.SERVICE
2
A136
Solenoid valve 6 failed
SOLENOID VALVE 6 ERROR
CONTACT TECH.SERVICE
2
A137
Solenoid valve 7 failed
SOLENOID VALVE 7 ERROR
CONTACT TECH.SERVICE
2
A140
Firmware update error
FW CLOUD UPDATE ERROR
1
A201
Pre-heating not executed within time-out (steam generator)
STEAM GENERATOR
RESISTOR OPEN CIRCUIT
CONTACT TECH.SERVICE
2
A202
Pre-heating not executed within time-out (tube bundle heating element)
HEATING BAND
OPEN CIRCUIT
CONTACT TECH.SERVICE
2
A250
1st pulse with vacuum not achieved within time out
1PV TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
2
A251
1st rise back up to atmospheric pressure not reached within the time-
out
ATM1 UPSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A252
1st pressure pulse not reached within the time-out
1PP UPSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A253
2nd pulse with vacuum not achieved within time out
2PV TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
2
A254
2nd rise back up to atmospheric pressure not reached within the time-
out
ATM2 UPSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A255
2nd pressure pulse not reached within the time-out
2PP UPSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A256
3rd pulse with vacuum not achieved within time-out
3PV TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
2
A257
3rd rise back up to atmospheric pressure not reached within the time-
out
ATM3 UPSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A258
3rd pressure pulse not reached within the time out
3PP UPSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A260
Chamber depressurisation not reached within time out
ATM3 DOWNSTROKE TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
2
A261
Chamber levelling not reached within time out
PRESSURE LEVELLING
TIMEOUT
CONTACT TECH.SERVICE
2
A262
Vacuum pulsation during drying not executed within time-out
PD PRESSURE
UPSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A353
1st drop to atmospheric pressure not completed within the time-out
ATM1 DOWNSTROKE TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
2
A356
2nd drop to atmospheric pressure not completed within the time-out
ATM2 DOWNSTROKE TIMEOUT
CHECK LOAD
CHECK CHAMBER FILTER
2
A360
Vacuum pulsation after maintenance step not executed within time-out
SPD PRESSURE
DOWNSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
A362
Chamber depressurisation during drying not reached within time-out
PD PRESSURE
DOWNSTROKE TIMEOUT
CONTACT TECH.SERVICE
2
1 = OK (warning)
2 = OK + door unl RESET