![SCHWIND Carriazo-Pendular Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/schwind/carriazo-pendular/carriazo-pendular_user-manual_1216777053.webp)
USER MANUAL
(Instruction for Use)
SW 2.9
Operation
Version 2.9.7 dated 2020-12-23
Page 53 of 75
_
Error No. Display 1
Display 2
Internal voltages
20
“Change battery!”
“Send system to service”.
21
“Clock battery!”
“Changed during next service (TSC)”.
22
“Missing V5V/A”
“Send system to service”.
23
“Missing V15V”
“Send system to service”.
24
“Missing Voltage -15V”
“Send system to service”.
25
“Missing V12V”
“Send system to service”.
26
“Missing V8V”
“Send system to service”.
27
“Missing V5V”
“Send system to service”.
Vacuum
30
“Repeat Vacuum Test!”
“Vacuum drops down slowly! Valve defective?”
31
“Can not create vacuum!”
“Block tubing and turn vacuum control clockwise”.
32
“Vac too high! Calibrate!”
“Turn vacuum control counter clockwise”.
33
“Vacuum
current
too
high!”
“System on room temp?” “Water in vac system?”
Pendular drive
40
“Motor current missing!”
“Check connections.” “Change motor or cable”
41
“Reverse
motion
blocked!”
“Release blockade by removing drive from head”.
Not displayed information:
Check connection drive/head.
Check condition of axis (wear out).
Check for foreign particles.
42
“Drive obstructed/dirty!”
“Clean worm wheel and its bearing”.
43
“Drive is blocked!”
“Clean worm wheel and its bearing”.
Blade drive
50
“Motor current missing”!
“Check connections”. “Change motor or cable”.
51
“Osc. motor overload!”
“Irrigate cutting head”. ”Cool down cutting head”.
Operation
60
“Vacuum loss detected!”
“Stop OP manually by foot pedal! No auto stop!”
61
“Vacuum restored!”
“Wait until reaching normal vacuum.”
62
“Vacuum loss - reaction”
“Reversing to home position”.
(Default setting)
Caution!
The
error no. 63
is optional only! (Depending on pre-setting, the head goes back or the treatment
will be continued. No announcement, only a beep signal appears).