USER MANUAL
(Instruction for Use)
SW 2.9
Operation
Version 2.9.7 dated 2020-12-23
Page 52 of 75
_
“Max vacuum OP start”
XXX mmHg
The operation can only be started when the vacuum is lower than
the Max OP start vacuum. As long the vacuum is higher the
system will produce a beep sound.
STD: 620 mmHg
MAX: 675 mmHg
“Min vacuum to start”
XXX mmHg
The operation can only be started when the vacuum is higher
than the Min OP start vacuum. As long the vacuum is lower the
system will produce a beep sound.
STD: 600 mmHg
MIN: 350 mmHg
Language“
XXXXXX
Choose among the languages English, German, Italian, French
and Spanish.
„Automatic switch off“
XX min
The system will be automatically switched off when it is out of
use.
STD: 90 min
MIN: 10 min
MAX: 90 min
„Firmware“
V2.9
System information.
“End of User menu”
“Select with arrow keys”
6.3.3
System Supervisor & Error Table
In case of an error the system will display the Error number, Date and time and a brief error
description (Display 1). After a few seconds the display will start alternating showing additional
information how to proceed (Display 2).
Error No. Display 1
Display 2
01
“System error”
“Send system to service”.
02
“Illegal EEPROM param”
“System param will be set to standard values”.
03
“Missing motor cable!”
“Connect motor cable”; “Change motor cable”.
04
“Parameter out of range!”
“Repeat calibration”; “Value not accepted!”
10
“Mains power off”
“Connect console to net!”
12
“Battery voltage low!”
“Check mains connection”.
13
“Battery is too low!”
“System will be switched off automatically”.
14
“Battery is too high”
“Disconnect from net” “Send system to service”