Schulze & Böhm GmbH, Hermülheimer Str. 10, 50321 Brühl Gebrauchsanweisung-Psori-Kamm-DE-EN-rev6.docx / 01.01.2016
S.19
In the case of failure:
Problem
Potential cause
Action
Device doesn’t work after
being dropped.
Damage of lamp or other
important parts.
Send the device to manufacturer.
Lamp can’t be switched on.
No power / defective power
outlet
Test the power outlet with a different
light. Use another power outlet if
necessary.
Lamp can’t be switched on.
Lamp is not correctly plugged in
the socket.
Take off the transparent cover and
check if the UV-lamp is plugged in
properly in the socket. If this doesn’t
work the UV-Lamp might be damaged.
Contact the manufacturer.
Therapy effect not as
expected.
Dosage too low.
Please get in touch with your treating
physician regarding the dose regimen.
Therapy effect not as
expected.
UV-Lampe too old,
Device older than 3 years
Send the device to manufacturer and
let the lamp being replaced.
Warranty
Warranty coverage is provided for 24 months from the date of set up and provides for free repair or
replacement provided the manufacturer is informed within 8 days of any apparent fault, and that the fault is
not the result of external damage, improper use, operational errors, or attempts on the part of the user to
repair the unit. Consumables such as the UV lamps are excluded from the warranty. Should a UV lamp fail
within the first 6 weeks of use, the lamp will be examined by the manufacturer and, if found to be defective
will be replaced by the manufacturer as a gesture of goodwill.
Liability on the part of the manufacturer
The manufacturer is only liable for the safety, reliability, and performance of the unit if the installation, any
additions, repairs, and other necessary measures are performed by the manufacturer or an individual
authorized by the manufacturer to perform these tasks; the unit is operated and handled by specialist
personnel or by the manufacturer or by authorized personnel trained by the manufacturer, and; the unit is
employed in compliance with the operating instructions and with the accessory equipment recommended
by the manufacturer.
Spare parts
UV-protection goggles
(Article-No. 4896)
Acrylic comb cover
(Artikel-Nr. 5102)
Countdown-Timer
(Article-No. 9885)
Acrylic comb row
(Article-No. 5103)
UVB-311 Lamp 9W
(Article-No. 9681)
UVA Lamp 9W
(Article-No. 1504)
Ballast 9W EU
(Article-No. 5071)