Dosing accuracy

Air bubbles in the system are the main cause of inaccuracy. It is absolutely decisive to keep the cylinder
free from air bubbles. The air bubbles are mainly adhering to the piston. The air bubbles can be removed
from the piston by knocking, and in the course of the subsequent dosing, or during rinsing, they will escape
outwards. Direct sun exposure and temperature fluctuations promote the formation of air bubbles in the
cylinder, since there is always (physically dissolved) air in the dosing or titration liquids.

If the system contains air bubbles which cannot be removed easily, one should predose a small volume,
e.g. 0.5 ml. Subsequently, use a double click of the grey key of the manual key button to reset the display
to zero. The burette is now ready for accurate dosing and titration.

7

Replacing the dosing unit (disassembly and installation)

As a rule, the need for replacing the dosing unit occurs only rarely. The dosing unit has to be re-
placed, if such a replacement becomes necessary as a result of a defect or of an inspection of the
TITRONIC

â

â

â

â

 

universal

  Piston Burette.

! Caution!  Please wear protective glasses!

Prior to starting working, screw off the hose from
the cylinder! 

!   

Chemicals might leak out!

The dosing unit is equipped with lateral ribs around its circumference, with one of these ribs being in dou-
ble design. This double rib serves as a mark for the correct placement of the dosing unit (

&

 Fig. 29,

item 2, or Fig. 9, item 2).

Important note: The disassembly of the dosing unit as such is only possible using a special tool (TZ 3630)
and should only be done by experienced persons. The sealing lip of the PTFE piston and the sealings of
the liquid system may be damaged or incorrectly positioned in the course of such a disassembly. As a re-
sult, titration liquid would leak in unwanted places.

7.1 Detaching and placement of the dosing unit

!

 

Caution, place a receptable under the titration tip!

 

!

Ü

Press the <FILL/F1> key for five seconds, and keep it depressed until the display shows <<change of
the dosing unit>>.
When the piston moves upwards, titration solution will be dosed out of the burette tip.

Ü

If <<unlock unit !!>> is displayed, the dosing unit has to be unlocked by rotating it by 90° to the left
(counter-clockwise Fig. 27).

Ü

Press <START>. The motor of the TITRONIC

â

â

â

â

 universal 

 Piston Burette will now automatically move

the dosing unit upwards. After the display shows <<remove unit / set on new unit>> remove the dosing
unit (

&

 Fig. 28). The process is completed.

Detaching the dosing unit :

Placement of the dosing unit:

Fig. 27  Unlocking

Fig. 28  Removal

Fig. 29  Placement

Fig. 30  Locking

Pos. 1

Pos. 2

Pos. 1

Pos. 2

Pos. 1

Pos. 2

M

Pos. 1

Pos. 2

Ü

Set on new (inspected, if necessary repaired) dosing unit vertically (

&

 Fig. 29), then press the <FILL>

key. The running motor will now pull the dosing unit downwards, while the display requests to align the
unit, „M“ mark to position 2. If any noise should become audible at this point, please stop the process
immediately by pressing the <START/STOP> key. The cause of this „rattling noise“ will be eliminated
by setting the dosing unit 

vertically

 on again!

Содержание TITRONIC universal

Страница 1: ...sanleitung Kolbenb rette TITRONIC universal Operating Instructions Piston Burette TITRONIC universal Mode d emploi Burette piston TITRONIC universal Manual de instrucciones Bureta de mbolo TITRONIC un...

Страница 2: ...iversal without changing the described properties Mode d emploi Page 49 72 Instructions importantes Pri re de lire et d observer attentivement le mode d emploi avant la premi re mise en marche de la B...

Страница 3: ...pplier s Certificate We certify that the equipment EN ISO 9001 part 4 10 4 Final inspection and testing and that the specified requirements for the product are met Certificat du fournisseur Nous certi...

Страница 4: ...TRONIC universal Piston Burette 33 4 1 Keys and contrast 33 4 2 Switch on image setting the filling time 33 4 3 Selecting the national language 34 4 4 Rinsing and initial filling 34 4 5 Filling 35 4 6...

Страница 5: ...26 Fig 1 TITRONIC universal Piston Burette with TM 96 Stirrer with manual key button with stand with electrode burette tip holder with burette tip TZ 1503 with bottle top TZ 2005...

Страница 6: ...contact with the titration and dosing solutions are made of chemically resistant mate rials Analysis accuracy is guaranteed by the high precision glass cylinder and the spindle for the piston feed Th...

Страница 7: ...ieht bereinstimmt mit den Normen und mit dem normativen Dokument We declare under our sole responsibility that the product Piston Burette TITRONIC universal to which this declaration relates is in con...

Страница 8: ...ading and 2 trailing decimal digits with a resolution of 0 01 ml Display Matrix LCD display 69 x 39 mm 64 x 128 pixels with background lighting contrast can be set using a knurled wheel Connections Ma...

Страница 9: ...ONIC universal Piston Burette from the place of work Examples for the assumption that a safe operation is no longer possible the package is damaged the TITRONIC universal Piston Burette shows visible...

Страница 10: ...he operating voltage of the TITRONIC universal Piston Burette matches the mains voltage The operating voltage is indicated on the specification plate on the bottom of the TITRONIC universal Piston Bur...

Страница 11: ...off Mains cable Interface 2 RS 232 C Connector for manual key button Interface 1 RS 232 C The transfer parameters for the RS 232 C interface can be selected according to RS 232 C interface chapter Att...

Страница 12: ...ered using the keyboard In dialogue mode the display shows which active keys can be pressed for further work The manual key button is used to start the START STOP and the FILL function and to output t...

Страница 13: ...ction of the cylinder contents Filling time F2 Setting the filling time System F3 Selection of the national language and setting of the serial parameters Menu Mode Selection of the functions Dosing ti...

Страница 14: ...al Piston Burette the display will show the respective basic menu of a function Here you use the FILL or the grey key of the manual key button to start the Filling function Upon completion of the manu...

Страница 15: ...you can reset the input to the last valid parameters by pressing the F4 key Fig 16 Fig 17 Dosing parameter Nominal vol 015 00ml Change digit Next digit OK F1 ESC F4 Dosing parameter Speed 40 00ml min...

Страница 16: ...wn including 2 nd stage high speed and keep it depressed The titration solution will be dosed as long as the key is kept depressed The display will show the volume and the selected rate stage with the...

Страница 17: ...ith the display of the TITRONIC universal Piston Burette Setting the parameters chapter The stored parameters remain active even after the burette is switched off 4 9 Incremental dosing Use the MODE k...

Страница 18: ...g with each other require the same transfer parameters The following transfer parameters are set ex factory Baud rate 4800 baud Data bits 7 Bit Stop bits 1 Parity No If you wish to change the data tra...

Страница 19: ...ton Burette This piston burette in turn will now check the address and will respond to this command just like the first burette Any information data strings arriving at the interface 2 of the TITRONIC...

Страница 20: ...ce Z 1 Titration Z 2 Dosing Z 3 Incremental dosing aaMO1 Set titration function aaY aaMO2 Set dosing function aaY aaMO3 Set incremental dosing function aaY aaRA Interchangeable unit inquiry of volume...

Страница 21: ...2 or Fig 9 item 2 Important note The disassembly of the dosing unit as such is only possible using a special tool TZ 3630 and should only be done by experienced persons The sealing lip of the PTFE pis...

Страница 22: ...the cylinder and in the hoses The disturbances to be anticipated are the greater the more the new solu tion differs from the previous type and concentration In the case of highly different solutions t...

Страница 23: ...rips after use Air bubbles in the system Set longer filling time When installing the dosing unit the piston rod is not pulled in correctly the device rattles Oblique application of piston rod Re apply...

Страница 24: ...30 If moisture is actually present one can conclude that the piston in the cylinder is no longer tight Check the hoses screwed connections and seals for visible damage contamination and leakage assem...

Страница 25: ...r repair please state the titration solution last used on the accompanying papers or on the TITRONIC universal Piston Burette itself Otherwise the service department will dispose of the damaged or rep...

Страница 26: ...uter PC with 25 channel plug 1 side 4 channel mini DIN plug 2 side 25 channel sub Miniature D plug TZ 3097 1 5 m data cable RS 232 C for computer PC with 9 channel plug 1 side 4 channel mini DIN plug...

Страница 27: ...48 Order numbers Fig 31 M3 X 10 mm TZ 3660 TZ 1503 TZ 3460 TZ 3025 TZ 2008 TZ 2005 TZ 3095 TITRONIC universal TZ 3665 TZ 3680...

Отзывы: