handylab multi 12
Operation
43
4.5
Calibration intervals (Int 3, Int 4, Int 5)
For each measured variable, a time interval is stored. When it has ex-
pired, you will be reminded to calibrate. After a calibration interval has
expired, the probe symbol of the relevant measured variable flashes. It
is still possible to measure. By calibrating the relevant probe, the func-
tion is reset and the interval starts anew.
The following calibration intervals are set in the factory:
Setting the calibration
interval
You can change each of these intervals (1 ... 999 days):
Measured parameter
Designation
Default setting
pH/ORP
Int 3
7 days
Dissolved oxygen
Int 4
14 days
Conductivity
Int 5
180 days
1
Switch off the measuring instrument.
2
Press the <M> key and hold it down.
3
Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.
4
Press the <RUN/ENTER> key repeatedly, until Int 3 ... 5 to-
gether with the required measured variable (pH, O
2
or
) ap-
pears on the display. Example:
5
Set the required interval until the next calibration with <
>
<
>.
6
Confirm with <RUN/ENTER>.
7
Switch to the measuring mode with <M>.
°
C
1
TP
ARng
mg/l
LoBat
Baud
Day.Month No.
Ident
Year
Sal
%
mbar
S
RCL
AR
Store
Auto
°
F
1/cm
/
K
%
TDS
UpH
Sal
mV/pH
S/
m
cm
cm
M
Lin
Tref25 Tref20
nLF
AutoCal DIN
AutoCal TEC
4
1
t 4
n
O
1
d
Time
Cal
Содержание handylab multi 12
Страница 72: ...What to do if handylab multi 12 72 ...
Страница 80: ...Index handylab multi 12 80 ...