handylab multi 12
Operation
53
Outputting the calibra-
tion protocol on the
interface
Sample printout:
1
Press the <RCL> key repeatedly until CAL SEr appears on the
display.
2
Press the <RUN/ENTER> key.
The calibration protocol for all measured variables is transmit-
ted to the interface. After the data transmission, the instrument
automatically switches to the last active measurement mode.
d
Store
Auto
Ident
S
1
LoBat
Tref25
nLF
/
K
%
Sal
°
AR
TP
C
M
m
%
mg/l
cm
S/
cm
cm
1/
°
F
ARng
Cal
Lin
Tref20
Sal
TDS
mbar
O
UpH
mV/pH
AutoCal DIN
AutoCal TEC
Time
Baud
Day.Month No.
Year
R
C A L
S E
RCL
CALIBRATION PROTOCOL
02.03.01
14:19
Device No.: 12345678
Calibration pH
Cal time: 01.03.01 / 15:20
Cal interval:
7d
AutoCal TEC
Tauto
Buffer 1
2.00
Buffer 2
4.01
Buffer 3
7.00 *
Buffer 4
10.01
C1
174.1 mV
25.0°C
C2
-133.3 mV
25.0°C
S1
-59.4 mV/pH
ASY1
-
4 mV
Probe:
+++
CALIBRATION 02
Cal time: 02.03.01 / 14:19
Cal interval:
14d
OxiCal
Tauto AR
Relative Slope:
0,88
Probe:
+++
Calibration Conductivity
Cal time: 14.12.00 / 11:37
Cal interval:
180d
Cal Std.:
0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const :
0.478
1/cm
Содержание handylab multi 12
Страница 72: ...What to do if handylab multi 12 72 ...
Страница 80: ...Index handylab multi 12 80 ...