Operation
handylab multi 12
20
4.2.2
Measuring the pH value
AutoRead AR
(Drift control)
The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.
For identical measurement conditions, the following criteria apply:
1
Perform the preparatory activities according to section 4.2.1.
2
Immerse the pH electrode in the test sample.
3
Press the <
> <
> keys until pH appears in the status display.
The pH value appears on the display.
4
When measuring without a connected temperature probe:
Options:
l
Determine the current temperature using a thermometer
and, while keeping the <RUN/ENTER> key depressed, en-
ter this temperature value with <
> <
>.
l
TP display indicator not displayed, socket for the second
probe is free:
Connect the second probe (Oxi or Cond) and immerse it in
the same sample. TP flashes, the temperature is automati-
cally measured using the second probe.
Tref20
Lin
Cal
8
1
S
Sal
O
U
%
mg/l
mV/pH
1/cmSal
AutoCal DIN
Tref25
nLF
AutoCal TEC
RCL
Auto Store
LoBat
ARng
AR
Time
Baud
Day.Month No.
Ident
Year
mbar
TDS
cm
M
°
F
/
K
%
S/
m
cm
8
4
2
9 9
6
°
C
TP
pH
Reproducibility
Response time
Better than 0.02
> 30 seconds
1
Call up the pH measuring mode with <
> <
>.
2
Activate the AutoRead function with <AR>.
The current measured value is frozen (hold function).
Содержание handylab multi 12