What to do if...
handylab multi 12
70
Measured value too low
Measured value too
high
6.3
Conductivity system messages
Error message
OFL
OFL
OFL
OFL
Error message
E3
E3
E3
E3
Cause
Remedy
– Insufficient flow
– Provide flow to the probe
Cause
Remedy
– High amount of dissolved
substances
– Correct solubility function
using the salinity equivalent
– Air bubbles bump on the
membrane with high velocity
– Avoid direct flow to the
membrane
– The carbon dioxide pressure is
too high (> 1 bar)
– Measuring not possible
Cause
Remedy
The measured value lies outside
the measuring range
– Measuring cell not connected
– Connect measuring cell
– Cable broken
– Replace measuring cell
Cause
Remedy
– Measuring cell contaminated
– Clean cell and replace it if
necessary
– Unsuitable calibration
solution
– Check calibration solutions