Operation
handylab multi 12
62
4.9
Reset
You can reset (initialize) the measurement parameters and the config-
uration parameters separately from one another.
Measurement
parameters
The following measuring parameters (pH/O
2
/Cond InI) can be reset to
the default condition:
pH
O
2
Conductivity
Note
The calibration data gets lost when the measuring parameters are re-
set. Recalibrate after performing a reset.
Configuration
parameters
The following configuration parameters (InI) are reset to the delivery
status:
Measuring mode
pH
Asymmetry
0 mV
Slope -59.16
mV
Measuring mode
D. O. concentration
Relative slope
1.00
Salinity (value)
0.0
Salinity (function)
Off
Measuring mode
Cell constant
0.475 cm
-1
(fixed)
Baud rate
4800
Display during the
pH calibration
Buffer nominal value
Interval 1
(automatically saved)
OFF
Interval 2
(for data transmission)
OFF