Calibration evaluation
After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis-
play.
Determining the
cell constant
You can determine the cell constant as follows:
Display
Cell constant [cm
-1
]
0.450 ... 0.500 cm
-1
E3
Eliminate the error according to
chapter 6 W
HAT
TO
DO
IF
...
Outside the range
0.450 ... 0.500 cm
-1
1
Press the <CAL> key repeatedly until the status display
CAL CELL appears.
2
Press the <RUN/ENTER> key. The CAL display appears, as
well as
– the current, calibrated cell constant (with sensor symbol on
the display) or
– the fixed cell constant 0.475 1/cm (without sensor symbol
on the display). In this case, the measurement parameters
are initialized (see section 4.9 R
ESET
).
3
Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.
4
Press the <RUN/ENTER>
key.
The AutoRead measurement to determine the cell constant
starts. The AR display indicator flashes until a stable signal is
reached. The cell constant determined is displayed. The mea-
suring instrument automatically stores the cell constants.
AR
AR
RCL
RCL
O
O
UpH
UpH
Sal
Sal
Store
Store
Auto
Auto
AutoCal DIN
AutoCal DIN
AutoCal TEC
AutoCal TEC
LoBat
LoBat
ARng
ARng
Tref20
Tref20
Lin
Lin
nLF
nLF
Sal
Sal
%
%
mg/l
mg/l
mV/pH
mV/pH
cm
cm
°
°
°
°
TP
TP
m
m
S/
S/
cm
cm
F
F
C
C
%
%
/
/
K
K
M
M
TDS
TDS
1
1
S
S
mbar
mbar
Time
Baud
Day.Month No.
Ident
Year
Time
Baud
Day.Month No.
Ident
Year
5
2
0
0
7
7
4
4
A
A
L
L
C
C
cm
cm
1/
1/
Tref25
Tref25
Cal
Cal
Содержание handylab multi 12
Страница 72: ...What to do if handylab multi 12 72 ...
Страница 80: ...Index handylab multi 12 80 ...