FRED easy
4.4
Semi-automatic defibrillation
Semi-Automatic Defibrillation
Step 1
Switching on and preparing the device
Step 2
Analysing the ECG signal
5. Briefly press the green button
(max. 1 second). A message prompts the user
to stay clear of the patient.
Patient hazard — The guidelines given in
4.1 Instructions and Safety Notes
must be
observed.
Fig. 4.7
Button to turn the device on/off
and to start analysis
1. Briefly press the green button (max. 1 second)
to switch on the device.
2. Assess the patient's condition (see
3. Connect the electrode cable to the device (see
4.3 Applying the adhesive elec-
4. Apply the defibrillation electrodes to the patient's chest (see
Briefly press the green button
(max. 1 second) to start the analysis. Otherwise, the de-
vice will switch off.
• The semi-automatic version of the
FRED easy®
can be configured so that the ECG
analysis is initiated automatically. In this case, the device prompts the user to con-
nect the electrodes and then automatically runs the analysis without any interven-
tion by the user.
• If the device detects ventricular fibrillation or ventricular tachycardia with a heart
rate exceeding 150 bpm,
follows; otherwise, continue with