Page 35
Defibrillation
4
User Guide
Applying the adhesive electrodes
4.3
A
rt.
no.
: 0-4
8
-00
13 Rev. m
FRED easy
4.3.2
Apply the adhesive electrodes and connect them to the device
Step 1
Open the electrode packaging
Fig. 4.2
Opening the electrode packaging
When using pre-connected electrodes, go directly to
Step 2
Connecting the electrode cable to the device
Fig. 4.3
Inserting the connector into the port
Risks for the user and the patient — The packaging of pre-connected electrodes
is welded to the electrode cable. Do not remove the packaging from the elec-
trode cable (risk of damaging the cable).
Pre-connected electrodes, which are already connected to the device, only need to
be applied to the patient’s chest (Step 3). In this case, Step 2 is not required. Elec-
trodes that are not pre-connected, need to be connected (Step 2) and applied (Step
3).