Page 15
Components and Operation
2
User Guide
Design
2.2
A
rt.
no.
: 0-4
8
-00
13 Rev. m
FRED easy
Available versions and options
FRED easy Life
Standard features for this device are the metronome and a non rechargeable Li/MnO2
battery. All other options listed above for semi-automatic operation are not available.
SCHILLER's service centre is able to configure various device functions by means of
a dedicated PC connection (see
).
Version
Available options for
automatic operation
Available options for
semi-automatic operation
SD card
• Metronome
• Rechargeable NiCd
battery
• ECG display
• Switchover to manual defibrillation
• Metronome
• Rechargeable NiCd battery
• Silent mode
Ethernet
• Metronome
• Rechargeable NiCd
battery
• ECG display
• Switchover to manual defibrillation
• Metronome
• Rechargeable NiCd battery
• Silent mode
Online
• Metronome
• ECG display
• Switchover to manual defibrillation
• Metronome
• Silent mode
Patients with implanted pacemakers
—
FRED easy®
features an electronic pacer
pulse suppression algorithm and therefore, pacemaker pulses are not taken into ac-
count for the analysis. Depending on the pacemaker model and on the position of the
electrodes, the compensation pulse following every pacer pulse may exceptionally
not work and be considered as a QRS complex. In this case, the analysis can be dis-
torted and inaccurate. It depends on the pacer pulse parameters whether or not the
compensation pulse is counted as a QRS complex.