65
MPS4232
Section 5: Software
EXAMPLE
SET VALZO 1
>
DEFAULT
0
RANGE
0 or 1
Set Field Calibration Points (FCAL)
COMMAND
FCAL
TYPE
Set/Get
DESCRIPTION
Sets the number of points and the min/max pressure for field calibration. Points will be evenly
spaced between the minimum and maximum and will include a 0.00 point.
SYNTAX
SET FCAL
<number of points> <minimum pressure> <maximum pressure>
RETURNS
Prompt when ready to accept a new command
EXAMPLE
SET FCAL 13 -5.00 5.00
>
DEFAULT
0
RANGE
These values are not range checked
Set Field Validation Points (FVAL)
COMMAND
FVAL
TYPE
Set/Get
DESCRIPTION
Sets the number of points and the min/max pressure for field validation. Points will be evenly
spaced between the minimum and maximum and will include a 0.00 point.
SYNTAX
SET FVAL
<number of points> <minimum pressure> <maximum pressure>
RETURNS
Prompt when ready to accept a new command
EXAMPLE
SET FVAL 15 -5.00 5.00
>
DEFAULT
0
RANGE
These values are not range checked
Содержание MPS4232
Страница 1: ...MPS4232 Miniature Pressure Scanner Hardware and Software Manual Software Version 1 00...
Страница 2: ......
Страница 3: ......
Страница 6: ...INTENTIONALLY LEFT BLANK...
Страница 12: ...6 MPS4232 Section 1 Specifications INTENTIONALLY LEFT BLANK...
Страница 15: ...9 MPS4232 Section 1 Introduction Section 1 Specifications MPS4232 INTENTIONALLY LEFT BLANK...
Страница 18: ...12 MPS4232 Section 2 Introduction INTENTIONALLY LEFT BLANK...
Страница 33: ...27 MPS4232 Section 4 Hardware INTENTIONALLY LEFT BLANK...
Страница 37: ...31 MPS4232 Section 5 Software INTENTIONALLY LEFT BLANK...
Страница 51: ...45 MPS4232 Section 5 Software...
Страница 81: ...75 MPS4232 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Страница 90: ...84 MPS4232 Appendix INTENTIONALLY LEFT BLANK...
Страница 94: ...88 MPS4232 Appendix D Software Revision Log Version 1 00 Initial Release...
Страница 95: ...MPS4232 89 Appendix INTENTIONALLY LEFT BLANK...