36
MPS4232
Section 5: Software
FUNCTION
BYTES
DATA TYPE
DESCRIPTION
Frame Time (ns)
4
U Integer
Frame Time in nanoseconds.
This value will either be a time relative to 0 (when PTP is disabled),
or an absolute PTP time.
Temperatures
16
Float
Array of 4 temperature values as read from the 4 RTDs.
Pressures
128
Float or
Integer
Array of 32 pressure values. The float or integer is selected based on
the units (float = EU, integer = RAW).
Table 5.2 - Padded 64 Channel Binary Data Packet
This binary packet is used when SET SIM == 64. When enabled, the packet size and structure will match a MPS4264,
64-channel binary packet, however will use padded data for channels not being used (pressure channels 33-64, tempera
-
tures 5-8). This allows the same packet structure to be used for MPS4200 32 and 64 channel modules in the same software.
This packet is a total of 348 bytes.
FUNCTION
BYTES
DATA TYPE
DESCRIPTION
Packet Type
4
Integer
A Hex
Packet Size
4
Integer
Size in bytes of this packet (348)
Frame Number
4
Integer
The current frame number if in the scan mode.
Serial Number
4
Integer
Serial Number of the module (SET SN)
Frame Rate
4
Float
Scanning rate in Hz
Valve Status
4
Integer
0 - Px Only (32 is valveless)
Units index
4
Integer
Units conversion index (add table)
Units conversion
factor
4
Float
Conversion factor from PSI to selected units
PTP Scan Start
Time (sec)
4
Integer
Scan start time in seconds
PTP Scan Start
Time (ns)
4
Integer
Scan start time in nanoseconds
External Trigger
Time
4
U Integer
External Trigger Time in us
Temperatures
32
Float
Array of 4 temperature values as read from the 4 RTDs. The
remaining bytes will be padded with “0” (5 - 8).
Pressures
256
Float or
Integer
Array of 64 pressure values. The float or integer is selected based on
the units (float = EU, integer = RAW).
Channels 1 - 32 will be real data. Channels 33-64 will be padded with
“0”.
Frame Time
4
Integer
Time the frame occurred (seconds)
Frame Time
4
Integer
Time the frame occurred (nanoseconds)
External Trigger
Time
4
Integer
Time the external trigger occurred (seconds)
External Trigger
Time
4
Integer
Time the external trigger occurred (nanoseconds)
Содержание MPS4232
Страница 1: ...MPS4232 Miniature Pressure Scanner Hardware and Software Manual Software Version 1 00...
Страница 2: ......
Страница 3: ......
Страница 6: ...INTENTIONALLY LEFT BLANK...
Страница 12: ...6 MPS4232 Section 1 Specifications INTENTIONALLY LEFT BLANK...
Страница 15: ...9 MPS4232 Section 1 Introduction Section 1 Specifications MPS4232 INTENTIONALLY LEFT BLANK...
Страница 18: ...12 MPS4232 Section 2 Introduction INTENTIONALLY LEFT BLANK...
Страница 33: ...27 MPS4232 Section 4 Hardware INTENTIONALLY LEFT BLANK...
Страница 37: ...31 MPS4232 Section 5 Software INTENTIONALLY LEFT BLANK...
Страница 51: ...45 MPS4232 Section 5 Software...
Страница 81: ...75 MPS4232 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Страница 90: ...84 MPS4232 Appendix INTENTIONALLY LEFT BLANK...
Страница 94: ...88 MPS4232 Appendix D Software Revision Log Version 1 00 Initial Release...
Страница 95: ...MPS4232 89 Appendix INTENTIONALLY LEFT BLANK...