DOMA019EN v8_10-2015– Shado
®
User Manual
Page 10 / 24
Security features of the electric models
Emergency stop and lowering
According to European standards all SCALEO Medical devices have a RED emergency stop
button which cuts the power off when set in the “ON “ position. To stop all the functions: simply
press on the red button until it is lowered. To put the functions back on: turn the button clockwise
until it is raised again.
The emergency lowering function allows the Shado
®
E to be swiftly lowered in case the patient
needs to be transferred out of the trolley or if control box is failing. With a ball-point pen press
strongly on the button marked "Emergency" to make the tray go down.
Be sure that your hand remains out of the way of the moving mechanical parts
during the lowering of the tray.
This function requires that the battery has not been removed and NOT fully
discharged
Contact SCALEO Medical if one of these elements has a dysfunction or if some of the
parts are damaged.
Charging and discharging the battery
The Shado
®
E is equipped with a 24 Vcc
battery/power pack and security devices which
protect the users of this equipment :
-
Emergency stop button. (RED button)
-
Voltage limitation with alarm in case of
battery unloading.
-
Emergency lowering button in case of
battery unloading.( marked EMERGENCY)
SCALEO Medical ships the Shado
®
E with
the emergency button placed in “ON”
position.
So before any use, make sure that the
emergency stop button is released to “OFF”
position.
When the battery is about to be completely discharged, a sound signal warns the user that the
battery should be charged. If the battery is not charged quickly, the unit turns itself off and the
raising/lowering functions are not active anymore.
The battery should be charged regularly. To do so, connect the main 24/Vcc transformer to the
Shado
®
connection current collector and then to the AC outlet.
A reserve allows the Shado
®
to be operated if the sound signal rings during the
showering of a patient enabling it to finish the transfer.
The battery must be recharged only with the chargers provided by SCALEO
Medical.
Charging a battery with a non adapted charger may cause over heating, damage
the different components and cause a fire hazard .