![Scaime eNodTouch Скачать руководство пользователя страница 9](http://html1.mh-extra.com/html/scaime/enodtouch/enodtouch_user-manual_1206005009.webp)
9
User manual SCAIME:
NU-eNodTouch-E-0413
2.2.6.
Maintenance menu
Menu Input
Action
Check Pattern
Check the drawing function. Confirms whether the liquid crystal display (LCD) displays correctly
Check Touch Panel
Touch panel check. Checks that the touched area lights up correctly.
Calibration
Correct an analog touch panel. Continue touching until the crosshatch moves or disappears.
Check com
Check the send and receive lines of network communication
3.
On-Line Functioning
3.1.
Startup screens
•
At Power-On, the connection screen is displayed while eNodTouch try to communicate with connected eNod4 device.
•
If the connection screen is displayed more than 5 seconds or if the communication error screen is displayed, consider
the connection cannot be established: check the RS485 wiring (§1.4) and check the eNod4 and eNodTouch
communication settings (§2.2.3).
Connection screen Communication error screen
•
When connection is well established with compatible eNod4 model, the following screen is displayed during 3 seconds.
eNod4 model and firmware version