User manual SCAIME:
NU-eNodTouch-E-0413
3.4.4.
Configuration Setup
Parameter
Possible Value
Description
Metrology Parameters
Stability criterion
None, 0.25 0.5, 1, 2d
Defines a stability interval, related to scale interval.
Zero at Startup
[Yes], [No]
Enable the Automatic Zeroing at power-on
Zero Tracking
[Yes], [No]
Enable the Zero tracking function
Digital Filtering parameters
A/D converter parameters
Rejection
[50Hz]-[60Hz]
The A/D converter have In-built adjustable low-pass filter with
frequency rejection of 50 Hz or 60 Hz harmonics.
Frequency
6 – 1600 Hz
Filtered measurement frequency is adjustable from 6.25 to 1600 Hz
Bessel low-pass filter parameters
Filter Order
0, 2, 3, 4
For the suppression of periodic vibrations, a Bessel low-pass digital
filter can be applied as an output of the A/D converter.
Minimum cut-off frequency value depend on the A/D converter rate:
0.10 Hz for 6.25 meas/s, 4 Hz for 400 meas/s
Cut-Off Frequency
0.1 to 200Hz
Notch Filter parameters
High & Low
Frequency
0.1 to 200Hz
Frequency band to attenuate with the digital Notch filter
Reset, factory setting
Allows resetting eNod4 or making eNod4 come back to factory settings