SBS-6000 19
Select the saved threshold value, click function key <-1><+1> to change
the serial No. and choose value. It will show in Gray color if it can not
configure. The display of select threshold value as follow:
TEST SETTING
2011-05-10 09:01:23
1. String Mode:
2. Cell Mode:
3. Conn Mode:
4. Save Mode:
5. Set Warning:
6. Set Test Type
:
Station
:
01
String
:
01
V Type
:
2 V- 0500AH
V Warning
:
02.10V
REF
:
001
2V:0500AH:0.483mΩ
7. Sample Test:
8. Zero Setting:
Enter
Esc
Srlect
Input
5.3.2
Cell setting
Functions description
Set the Cell setting parameters like: Battery type, Low pressure alarm
value and reference value. Battery type includes battery voltage and
capacity. Low pressure alarm value is the alarm threshold value of the
measured battery, the instruments will alarm when the measured
voltage lower than the threshold one.
How to use
Comparison threshold value setting include: Manual Input directly or
Select threshold value. Select threshold means choose the Built-in set
value. The Build-in values can get directly from the sampling battery
testing or the configured download software. Details please take the
chapter “Sampling Battery” for reference. Display of Cell setting as
follow: