SBS-6000 39
How to use
Click String testing
icon that may enter into the string testing
interface as below:
Range
:
4 mΩ
2 . 1 1 2 (V)
0 . 5 37 (mΩ)
T
25.0 (
℃
)
2011-05-10 09:01:23
Manual
String
:
01
Cell No.002
Station
:
01
REF : <0.483mΩ> 1.11 REF
V
R
STRING TEST
Save
Esc
Cancel
Range
Connection method can refer to “Test Functions” Chapter
Before string testing you have to setup parameter of test strings,
detailed operation please go for the “String Testing Setup” Chapter.
When complete the setup if alarm threshold need changed again or
test type is invalid to current test strings. Each string can save 430 items
battery, in the condition of manual test, save data has to wait until test
data to be stabilized.
5.4.2
Cell Testing
Functions description
When tester performs Cell testing, that can test battery voltage,