SBS-6000 37
auto test and save data automatically.
Group, Station,
Battery
number
Display current group, station, battery number
three types of test status and display them in
different ways.
Test display
Display test results, string testing and Cell testing
display: voltage, impedance and temperature, Conn
display impedance and temperature.
Range
Display impedance test use the current gear, if the
test type is conductivity that would not display gear.
Functional
keys
Four functions correspondence to F1-F4 keys, in
difference functions has different interfaces.
Reference
value and
results
compare
(Test types is
resistance)
“REF: 0.483 m
Ω
” stands for reference value, 0.483
m
Ω
, ”1.11REF” stands for tested impedance value
is 1.11 times of reference value. When impedance
and voltage test value marked as
means
test values does not exceed the alarmed threshold.
Otherwise, tested value marked as
that
means tested battery voltage is exceed alarms.
Tested impedance marked as
means test
results are exceeded replaced value of configured
threshold. When impedance marked as
that means compare result is greater than warning
value of setup alarm threshold.