24
Error messages
An error code will appear on the display if any error occurs in the unit.
Code Description
Action
E02
The measured value is below the lower
limit of the display range.
Use the unit within the measurement range.
E03
The measured value is above the upper
limit of the display range.
E04
Calculation error. The calculation is
performed with the measured value in an
error state.
Confirm the probe is properly connected and
use the unit within the measurement range.
E07
The measurement environment is below
the lower limit of the operating
temperature range.
Use the unit within the operating temperature
range.
E08
The measurement environment is above
the upper limit of the operating
temperature range.
E09
The setting of the preset start/stop date
and time is abnormal.
The preset start date and time must be set to
later than the current date and time.
The preset stop date and time must be set to
later than the preset start date and time.
E10
Logging or memo cannot be performed
due to lack of space on the SD card.
Turn off the power, free up space on the SD
card, release its write-protect (lock), and then
reinstall it.
E11
The SD card cannot be recognized.
(during preset standby)
Turn off the power and reinstall the SD card.
Note: Set the preset logging again if
necessary.
E12
The SD card cannot be properly read.
- The SD card may not be properly
connected. Turn off the power, check that
there is no dust or dirt on the SD card
terminals, and then reinstall it.
- The SD card may not supported by the unit.
See "Installing the SD card".
E13
Logging or memo cannot be performed
due to low battery level.
Replace the batteries with new ones.
Содержание 8191-00
Страница 28: ......