System Drainage for Long Term Storage
Draining the system will prevent sediment build up or freezing. This section will
provide you with the procedures and methods needed to successfully drain the
system. This equipment is designed to provide years of service if maintained
properly.
WARNING!
DO NOT store system where it may be subjected to below freezing
temperatures without proper draining.
To drain the system use the following procedure:
1.
Turn off main power.
2.
Disconnect the air supply.
3.
Shutoff and disconnect the water supply.
4.
Bleed off the pressure. Slowly open the internal by-pass valve.
5.
Remove the water filter canister and drain water from it. Replace the drained
canister.
6.
Connect an air supply to the water supply port. Turn on air supply.
7.
Close the internal by-pass valve when the system is completely drained.