Series 3100
PP3123-03-0101
Automatic Burst Test Systems
User’s Guide
SATEC PIPE & PRESSURE
26
O
PERATION
Filling a Specimen
1.
Turn main power circuit breaker ON.
WARNING!
Specimen end caps must be of sufficient rating for the test pressures,
temperature and the specimen.
WARNING!
Check the thrust capacity of the test fixture, specimen and the enclosures
before pressurizing your specimen.
2.
Install specimen end caps on the specimen.
3.
Hook up the specimen in the burst tank.
4.
Open the customer supplied purge valve.
5.
Make sure the
STOP TEST
button is disengaged (in the out position).
6.
Open the
SPECIMEN FILL
valve.
7.
Close the purge valve.
8.
When all of the air in the sample has been displaced by water, close the
SPECIMEN FILL
valve.
9.
Close the tank lid. Be sure the
ENCLOSURE OPEN
indicator light is out.