Parameters
One parameter is used in the Eurotherm controller; the “
LogH
” parameter.
“
LogH
” measures the maximum pressure reached in the specimen during a test.
At the end of each test, “
LogH
” can be read by pressing the Scroll button.
“
LogH
” does not automatically reset itself at the beginning of a new test;
therefore it is necessary to reset it manually:
1.
Press the Scroll button until “
rES.L
” parameter is displayed. This is the
parameter that resets “
LogH
”. “
YES
” should be displayed below “
rES.L
”. If
“
YES
” is not displayed, use the Up/Down buttons to change this to “
YES
”.
2.
Press Page and Scroll simultaneously to return to the
Home
display.
Alarms
Alarms are flashed as messages in the lower readout of the
Home
display. A new
alarm is displayed as a double flash followed by a pause. If there is more than one
alarm condition, the display cycles through all the relevant alarm messages. There
are three alarms that the Eurotherm is configured for:
•
1FSH – timer start pressure, typically set at 75 psi
•
3FSH – specimen failure limit, typically set at 250 psi
•
2FSL – specimen failure limit + 10 psi, typically set at 260 psi
As the burst system pressurizes a specimen, each alarm message will flash in the
lower readout as that pressure is reached throughout the test. These alarm settings
should not be changed. Contact INSTRON-SATEC SYSTEMS if you feel that
these settings should be changed.